The PE25-C general purpose pyroelectric energy sensor has a 24 mm aperture. It can measure energies from 8 µJ up to 10 J. It can operate at repetition rates up to 10 kHz and covers the spectral range from 150 to 3000 nm.
High rep rate metallic absorber with a 150-3000 nm spectral range
Additional CAD file downloads are not available for this product.
pe25-c-pyroelectric-sensor - Drawings
Additional drawings are not available for this product.
PE-C Sensors with StarLab
Support for PE-C \ PD10-C sensors with Vega, Nova II, Juno, Juno+, Juno-RS, EA-1, StarLite and StarBright in StarLab application is now available. Follow step-by-step instructions to upgrade the embedded software in the PE-C \ PD10-C sensor with the StarLab application
When working with other meters and interfaces it is not necessary to update the embedded software of the PE-C \ PD10-C.
Note: PE-C \ PD10-C sensors can't be upgraded via EA-1 interface..
Specifications
Product Name
PE25-C
Aperture Size
Ø24 mm
Absorber Type
Metallic
Spectral Range
150-3000 nm
Minimum Pulse Energy
8 µJ
Maximum Pulse Energy
10 J
Maximum Frequency
10,000 Hz
Maximum Average Power
15 W
Maximum Average Power Density
20 W/cm²
Maximum Pulse Width
5 ms
Damage Threshold (100 ns)
0.1 J/cm²
Damage Threshold (2 ms)
6 J/cm²
Cable Length
1.5 m
Dimensions
Ø62 x 21 mm
CE Compliance
Yes
UKCA Compliance
Yes
China RoHS Compliance
Yes
Features
Measuring Laser Energy with a Pyroelectric Sensor
In this short “Basics” video, we review in general the use of pyro-electric sensors for measuring laser pulse energies.
Premium Energy Sensors – Even for UV
Meet Ophir’s high performance energy sensors for pulsed UV lasers (and not only for UV…). These sensors are the perfect solution for pulsed lasers at 193 and 248 nm.
Energy Sensor Response & Integration Time
There seems to be a good deal of confusion when it comes to the terms “Response time” and “Integration time” of energy sensors. In this video we will clarify the meaning of these terms, as they apply to Ophir’s pyro-electric “Smart Sensors”.
How to Detect Missing Laser Pulses with a Pyroelectric Sensor
This video will show you how to detect and record any missing pulses in your pulsed laser beam application.
Additional Error with Frequency in a Pyroelectric Sensor
In many laser energy sensor specifications, there is an item called "Additional error with frequency". Learn what that means, and why and when it happens, in this video.
The Pyro-C sensors have a "user threshold" feature allowing the user to adjust the measurement threshold in noisy environments. Increasing the threshold will prevent triggering on noisy signals and allow accurate measurment of energy and frequency, as long as the laser pulses are larger than the noise.
The trigger level can be adjusted up to 25% of full scale, however operation depends on the pulse width setting. For pulse width settings below ~0.25ms, the minimum energy that can be measured accurately is approximately 40% above the user threshold setting. Pulses below this energy level will trigger the sensor down to the user threshold level, but accuracy is compromised.
For pulse width settings above ~0.25ms, accuracy is good all the way down to the threshold. If the laser pulse width is less than 1/2 the setting, the minimum energy corresponds to the setting. However, with longer laser pulse widths, the minimum energy will be higher, rising to approximately twice the user threshold level when the laser pulse width is equal to the sensor pulse width setting.
It is recommended always to set the user threshold to the minimum possible setting to retain best energy accuracy in any given situation. See the user manual for further information on how to use the user threshold.
All Ophir pyroelectric sensors can measure average power with Ophir Power and Energy Meters. The instrument measures the number of pulses each second and multiplies the energy reading by the pulse rate. If the pulse rate is constant, then the accuracy of power measurement will be the same as the energy accuracy since the pulse rate measurement is very accurate.
First, clean the absorber surface with a tissue, using Umicore #2 Substrate Cleaner, acetone or methanol. Then dry the surface with another tissue. Please note that a few absorbers (Pyro-BB, 10K-W, 15K-W, 16K-W and 30K-W) cannot be cleaned with this method. Instead, simply blow off the dust with clean air or nitrogen. Don't touch these absorbers. Also, HE sensors (such as the 30(150)A-HE-17) should not be cleaned with acetone.
Note: These suggestions are made without guarantee. The cleaning process may result in scratching or staining of the surface in some cases and may also change the calibration.
The Power and Energy Meters simply decides it is time for a sample and takes the next pulse that comes after that time, e.g. if it samples at 400 Hz, then every 1/400th of a second it is ready to take the next pulse that comes along.
Yes, with certain limitations. Here are the points to be aware of:
Vega, Nova II, StarLite, StarBright meters and Juno PC interface: Full support of all features
All other instruments (Nova/Orion and LaserStar meters, as well as USBI, Pulsar, and Quasar PC interfaces): Support the Pyro-C sensors, except for the following features: Only 2 of the 5 pulse width settings are available.
User selectable threshold is not available.
In addition to the above: When using a Pyro-C sensor with the Nova (or Orion) meter, the "Nova PE-C Adapter" (Ophir p/n 7Z08272) is required.
The catalog specification states the maximum power a sensor can be used with and without the heat sink. The purpose of the heat sink is to keep the sensor temperature below the maximum permitted at higher average powers. If you use the sensor for a short time only, on the order of 1-2 minutes at a time, you should be able to measure up to the higher power given in the spec even without the heat sink.
The old pyro sensors and the newer PE-C sensors are almost identical. The differences between them are as follows:
More compact
User Threshold – minimum energy threshold (below which the sensor will not trigger) can be selected according to users' needs
Measures longer pulses (up to 20ms depending on model)
Has up to 5 pulse width settings as opposed to only 2 pulse width settings
Disadvantages: Smaller size and therefore:
May need a heat sink (P/N 7Z08267) in order to stand up to higher average powers
May need a mechanical size adapter (P/N 7Z08273) if it must fit into an existing mechanical jig designed for the older models
Meters and Software Support: StarLite, Juno, Vega, & Nova II fully support the Pyro-C series. Laserstar, Pulsar, USBI, Quasar, and Nova / Orion with adapter* partially support the Pyro-C series:
Only 2 of the 5 pulse width settings are available
Lowest measureable energy cannot be selected (no User Threshold).
StarLab software supports both Pyro-C and older pyro series.
*Note: The PE-C series will only operate with Nova / Orion meters with an additional adapter Ophir P/N 7Z08272 (see details in Ophir website).
Wavelength Setting Names: If you have your own software for communicating with the sensor, it may be important to note that for some models, the names of the wavelength settings are a bit different between the old pyro and the new PE-C, even though they mean exactly the same thing.
For example, with diffuser OUT, the settings in the PE50BB-DIF-V2 are called “<.8u” (i.e. visible, represented by a calibration point at 532nm that covers the full visible range), and “106” (i.e. 1064nm), while in the PE50BB-DIF-C these same settings are called “532” (i.e. 532nm, the calibration point for the visible) and “1064”.
In theory, if a beam is completely parallel and fits within the aperture of a sensor, then it should make no difference at all what the distance is. It will be the same number of photons (ignoring absorption by the air, which is negligible except in the UV below 250nm). If, nevertheless, you do see such a distance dependence, there could be one of the following effects happening:
If you are using a thermal type power sensor, you might actually be measuring heat from the laser itself. When very close to the laser, the thermal sensor might be “feeling” the laser’s own heat. That would not, however, continue to have an effect at more than a few cm distance unless the light source is weak and the heat source is strong.
Beam geometry – The beam may not be parallel and may be diverging. Often, the lower intensity wings of the beam have greater divergence rate than the main portion of the beam. These may be missing the sensor's aperture as the distance increases. To check that you'd need to use a profiler, or perhaps a BeamTrack PPS (Power/Position/Size) sensor.
If you are measuring pulse energies with a diffuser-based pyroelectric sensor: Some users find that when they start with the sensor right up close to the laser and move it away, the readings drop sharply (typically by some 6%) over the first few cm. This is likely caused by multiple reflections between the diffuser and the laser device, which at the closest distance might be causing an incorrectly high reading. You should back off from the source by at least some 5cm, more if the beam is not too divergent.
Needless to say, it’s also important to be sure to have a steady setup. A sensor held by hand could easily be moved around involuntarily, which could cause partial or complete missing of the sensor’s aperture at increasing distance, particularly for an invisible beam.
When logging energy measurements on a PC with the StarLab software from a Pyro sensor via either a Nova-II, or Vega, or a USB enabled StarLite meter, the timestamp for each Energy pulse measured in the log is provided entirely by the clock on the PC which has millisecond resolution. (Note: Because a timestamp provided by a multitasking Windows PC is not from a true real time system, there could be instances where the timestamp is not well synced with the actual energy pulse measurement in the log, depending on how ‘burdened’ the computer was at any particular moment.)
When logging energy measurements on a PC with the StarLab software from a Pyro sensor via either a StarBright, Juno or Pulsar, each of these meters provides a precise microsecond resolution timestamp from their on-board clock.
This timestamp is synced to the Energy measurement and the data is written together in the log. The precise on-board clock in the StarBright, Juno or the Pulsar is used here to determine the time differences between measurements - rather than the PC clock which is used here just to set the initial baseline time of the log. This is the best method to log Energy if timing of pulses is critical.
As opposed to Pyro Energy measurements, when logging Power measurements on a PC via StarLab with either Photodiode or Thermopile sensors, where fast measurements are not required anyway, the log timestamp is provided entirely by the millisecond resolution clock on the PC when connected to any of our meters.
The Pyro-C sensors have a "user threshold" feature allowing the user to adjust the measurement threshold in noisy environments. Increasing the threshold will prevent triggering on noisy signals and allow accurate measurment of energy and frequency, as long as the laser pulses are larger than the noise.
The trigger level can be adjusted up to 25% of full scale, however operation depends on the pulse width setting. For pulse width settings below ~0.25ms, the minimum energy that can be measured accurately is approximately 40% above the user threshold setting. Pulses below this energy level will trigger the sensor down to the user threshold level, but accuracy is compromised.
For pulse width settings above ~0.25ms, accuracy is good all the way down to the threshold. If the laser pulse width is less than 1/2 the setting, the minimum energy corresponds to the setting. However, with longer laser pulse widths, the minimum energy will be higher, rising to approximately twice the user threshold level when the laser pulse width is equal to the sensor pulse width setting.
It is recommended always to set the user threshold to the minimum possible setting to retain best energy accuracy in any given situation. See the user manual for further information on how to use the user threshold.
All Ophir pyroelectric sensors can measure average power with Ophir Power and Energy Meters. The instrument measures the number of pulses each second and multiplies the energy reading by the pulse rate. If the pulse rate is constant, then the accuracy of power measurement will be the same as the energy accuracy since the pulse rate measurement is very accurate.
First, clean the absorber surface with a tissue, using Umicore #2 Substrate Cleaner, acetone or methanol. Then dry the surface with another tissue. Please note that a few absorbers (Pyro-BB, 10K-W, 15K-W, 16K-W and 30K-W) cannot be cleaned with this method. Instead, simply blow off the dust with clean air or nitrogen. Don't touch these absorbers. Also, HE sensors (such as the 30(150)A-HE-17) should not be cleaned with acetone.
Note: These suggestions are made without guarantee. The cleaning process may result in scratching or staining of the surface in some cases and may also change the calibration.
The Power and Energy Meters simply decides it is time for a sample and takes the next pulse that comes after that time, e.g. if it samples at 400 Hz, then every 1/400th of a second it is ready to take the next pulse that comes along.
Yes, with certain limitations. Here are the points to be aware of:
Vega, Nova II, StarLite, StarBright meters and Juno PC interface: Full support of all features
All other instruments (Nova/Orion and LaserStar meters, as well as USBI, Pulsar, and Quasar PC interfaces): Support the Pyro-C sensors, except for the following features: Only 2 of the 5 pulse width settings are available.
User selectable threshold is not available.
In addition to the above: When using a Pyro-C sensor with the Nova (or Orion) meter, the "Nova PE-C Adapter" (Ophir p/n 7Z08272) is required.
The catalog specification states the maximum power a sensor can be used with and without the heat sink. The purpose of the heat sink is to keep the sensor temperature below the maximum permitted at higher average powers. If you use the sensor for a short time only, on the order of 1-2 minutes at a time, you should be able to measure up to the higher power given in the spec even without the heat sink.
The old pyro sensors and the newer PE-C sensors are almost identical. The differences between them are as follows:
More compact
User Threshold – minimum energy threshold (below which the sensor will not trigger) can be selected according to users' needs
Measures longer pulses (up to 20ms depending on model)
Has up to 5 pulse width settings as opposed to only 2 pulse width settings
Disadvantages: Smaller size and therefore:
May need a heat sink (P/N 7Z08267) in order to stand up to higher average powers
May need a mechanical size adapter (P/N 7Z08273) if it must fit into an existing mechanical jig designed for the older models
Meters and Software Support: StarLite, Juno, Vega, & Nova II fully support the Pyro-C series. Laserstar, Pulsar, USBI, Quasar, and Nova / Orion with adapter* partially support the Pyro-C series:
Only 2 of the 5 pulse width settings are available
Lowest measureable energy cannot be selected (no User Threshold).
StarLab software supports both Pyro-C and older pyro series.
*Note: The PE-C series will only operate with Nova / Orion meters with an additional adapter Ophir P/N 7Z08272 (see details in Ophir website).
Wavelength Setting Names: If you have your own software for communicating with the sensor, it may be important to note that for some models, the names of the wavelength settings are a bit different between the old pyro and the new PE-C, even though they mean exactly the same thing.
For example, with diffuser OUT, the settings in the PE50BB-DIF-V2 are called “<.8u” (i.e. visible, represented by a calibration point at 532nm that covers the full visible range), and “106” (i.e. 1064nm), while in the PE50BB-DIF-C these same settings are called “532” (i.e. 532nm, the calibration point for the visible) and “1064”.
In theory, if a beam is completely parallel and fits within the aperture of a sensor, then it should make no difference at all what the distance is. It will be the same number of photons (ignoring absorption by the air, which is negligible except in the UV below 250nm). If, nevertheless, you do see such a distance dependence, there could be one of the following effects happening:
If you are using a thermal type power sensor, you might actually be measuring heat from the laser itself. When very close to the laser, the thermal sensor might be “feeling” the laser’s own heat. That would not, however, continue to have an effect at more than a few cm distance unless the light source is weak and the heat source is strong.
Beam geometry – The beam may not be parallel and may be diverging. Often, the lower intensity wings of the beam have greater divergence rate than the main portion of the beam. These may be missing the sensor's aperture as the distance increases. To check that you'd need to use a profiler, or perhaps a BeamTrack PPS (Power/Position/Size) sensor.
If you are measuring pulse energies with a diffuser-based pyroelectric sensor: Some users find that when they start with the sensor right up close to the laser and move it away, the readings drop sharply (typically by some 6%) over the first few cm. This is likely caused by multiple reflections between the diffuser and the laser device, which at the closest distance might be causing an incorrectly high reading. You should back off from the source by at least some 5cm, more if the beam is not too divergent.
Needless to say, it’s also important to be sure to have a steady setup. A sensor held by hand could easily be moved around involuntarily, which could cause partial or complete missing of the sensor’s aperture at increasing distance, particularly for an invisible beam.
When logging energy measurements on a PC with the StarLab software from a Pyro sensor via either a Nova-II, or Vega, or a USB enabled StarLite meter, the timestamp for each Energy pulse measured in the log is provided entirely by the clock on the PC which has millisecond resolution. (Note: Because a timestamp provided by a multitasking Windows PC is not from a true real time system, there could be instances where the timestamp is not well synced with the actual energy pulse measurement in the log, depending on how ‘burdened’ the computer was at any particular moment.)
When logging energy measurements on a PC with the StarLab software from a Pyro sensor via either a StarBright, Juno or Pulsar, each of these meters provides a precise microsecond resolution timestamp from their on-board clock.
This timestamp is synced to the Energy measurement and the data is written together in the log. The precise on-board clock in the StarBright, Juno or the Pulsar is used here to determine the time differences between measurements - rather than the PC clock which is used here just to set the initial baseline time of the log. This is the best method to log Energy if timing of pulses is critical.
As opposed to Pyro Energy measurements, when logging Power measurements on a PC via StarLab with either Photodiode or Thermopile sensors, where fast measurements are not required anyway, the log timestamp is provided entirely by the millisecond resolution clock on the PC when connected to any of our meters.
Accessories
Heat Sink
Screws onto rear of PE-C series sensors allowing working at over 50% higher average powers
These adapters allow for power measurement of connectorized fiber-optic cables. The sensor may need an additional mounting bracket to connect to these fiber adapters.
Compare
Description
Compatibility
Drawings, CAD & Specs
Avail.
Price
7Z08227SC Fiber Connector Adapter, Power and Energy Sensors
UNIVERSAL
In Stock
$302.00
7Z08226ST Fiber Connector Adapter, Power and Energy Sensors
UNIVERSAL
In Stock
$264.00
7Z08229FC Fiber Connector Adapter, Power and Energy Sensors
UNIVERSAL
In Stock
$153.00
1G01236ASMA Fiber Connector Adapter, Power and Energy Sensors
UNIVERSAL
In Stock
$74.00
Showing 1 to 4 of 4 entries
Fiber Adapter Mounting Bracket
A mounting bracket is needed to connect most power and energy sensors to a fiber adapter (SC, ST, FC or SMA). This bracket can be used for pyroelectric energy sensors in the PE9, PE10 and PE25 series. It is not compatible with diffuser-equipped sensors.
The newer PE-C series sensors have a Ø62 mm diameter. The older PE series sensors have a Ø85 mm diameter. This adapter allows using the PE-C type sensors in jigs and setups that were originally designed for PE sensors.
Beam Splitter Assembly to measure pulsed laser sources too energetic for direct measurement. The reading with the Beam Splitter can be calibrated by setting the laser to a lower energy that will not damage the sensor and then taking a measurement with the beam splitter and without and taking the ratio.
Compare
Description
Compatibility
Drawings, CAD & Specs
Avail.
Price
7Z17001Beam Splitter Assembly, Pyroelectric Energy Sensor
UNIVERSAL
In Stock
$802.00
Showing 1 to 1 of 1 entries
Damage Threshold Test Plates (Metallic)
Test plates with same absorber coating as the sensor. For testing that laser beam is not above damage threshold (1 such plate is included with sensor package).
Order a fixed extension cable along with sensor to receive a different cable than the standard 1.5 m cable.
Compare
Description
Compatibility
Drawings, CAD & Specs
Avail.
Price
7E01174AFixed extension cable for Pyro Energy Sensors, 3 m Length, Connect to Power Meter
UNIVERSAL
In Stock
$86.00
7E01174BFixed extension cable for Pyro Energy Sensors, 5 m Length, Connect to Power Meter
UNIVERSAL
5 Weeks
$112.00
7E01174CFixed extension cable for Pyro Energy Sensors, 10 m Length, Connect to Power Meter
UNIVERSAL
In Stock
$167.00
Showing 1 to 3 of 3 entries
Modular extension cables for Pyro Energy Sensors
The extension cables can be ordered separately from the sensors.
Compare
Description
Compatibility
Drawings, CAD & Specs
Avail.
Price
7E01198Extension cable for Pyro Energy Sensors, 1.5 m Length, D15M-D15F
UNIVERSAL
In Stock
$68.00
7E01198AExtension cable for Pyro Energy Sensors, 3.5 m Length, D15M-D15F
UNIVERSAL
In Stock
$89.00
7E01198BExtension cable for Pyro Energy Sensors, 8.5 m Length, D15M-D15F
UNIVERSAL
In Stock
$134.00
Showing 1 to 3 of 3 entries
Extended Warranty for Sensor
Customers that purchase the above items also consider the following items. Ophir-Spiricon meters and sensors include a standard manufacturers warranty for one year. Add a one year Extended Warranty to your meter or sensor, which includes one recalibration.
In order to provide better service and products, please provide the following brief information. Any future resource requests will be automatically available.
Password Reset
Enter your email address below to reset your account password.
Password Reset
Email Verification Required
Cart Items Updated
Remove Product
Remove this product from your comparison list?
Check Order Status
Provide an order number and postal code to check the status of an order or download an invoice for an order that has shipped. Login to view your complete order history.
Sign In Required
To access this and other valuable technical resources, please sign in or register for a new online account.